Voltage
Charging voltage range: 0V ~ 4.3V (Remarks: If the customer needs charging measurement range 0 ~ 5V, it needs to be customized)
Discharge voltage range: 2.5V~5V
Accuracy: ±(0.1% of reading + 0.1% of range)
Electric current
Output range: ±3000mA
Accuracy: ±(0.1% of reading + 0.1% of range)
Constant voltage cut-off current: 15mA minimum
Time
Inspection time: Real time<2s
Timing range: 0 ~ 9999 minutes
Data logging interval: Time ≥20s
Charge
Charging mode: Constant current charging, nickel metal hydride charging, lithium battery charging
Cut-off condition: Voltage, time, cut-off current, -ΔV
Discharge
Discharge mode: Constant current discharge
Cut-off condition: Voltage, time
Cycle
Cycle range: 1 ~ 999 times
Loop nesting: With nested loop function, maximum support 3 levels of nesting
Single cycle working steps: 254
Protection
Software protection: Power-down data protection
Safety protection conditions can be set, and overvoltage protection, undervoltage protection, and protection voltage can be set